Add to Cart
Recessed Through Wall Hospital Sterilizer Equipment With Pre-vacuum And Post Vacuum Phase
Specifications
Model No | Doors | Heating Type | Power Supply | Volume | Internal Chamber Dimension | Overall Dimension |
(Liters) | (w x h x d) | (w x h x d) | ||||
MD-0.8Y | Sliding door | External steam supply | AC380V, 50Hz, 3P/N/E | 800 | 680*1180*1000 | 2200*2000*1410 |
Outside facilities needed | ||||||
Electric power:AC380,50HZ,3 phase/N/E | ||||||
Steam feeding: Saturated steam from external steam pipes 0.4-0.7Mpa | ||||||
Water: Softened water, hardness <50ppm, pH Neutral, 0.15-0.3Mpa | ||||||
Compressed air: Oil free air, 0.5-0.8Mpa |
Program Parameter
Cycle name | Further details | Application |
Instruments/Tools | 3 times Vacuum, expose temperature 132℃, expose time 8 min, drying time 5 min. | Solid materials to be sterilized: ●Packaged instruments ●Empty glass containers ●Empty metal containers Porous materials to be sterilized: ●Rubber articles |
Clothing | 3 times Vacuum, expose temperature 121℃, expose time 15 min, drying time 25 min. | Materials to be sterilized from which air can’t be easily displaced ●Textile |
User Programmable | All preset parameters can be reset; Multiple customized programs can be set and stored by user. | Customized |
Test programs | ||
B-D Test | 3 times Vacuum, expose temperature 134℃, expose time 4 min, drying time 3 min. | Test program |
Vacuum Leakage Test | Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect. | Test program |
Description
This series hospital sterilizer operate with saturated steam as the sterilizing agent with a
temperature range of 105 °C (221°F) to 134 °C (272.2 °F) and a working chamber pressure
that meets Supervision Regulation on Safety Technology for Fixed Type Pressure Vessel.
With years of research and customer feedback we have fine-tuned our formula for perfectly
optimized cycles with the precise combination of pulses, timing, temperature control, speed
and drying.
Applications
Bedspreads, linens, laundry, medical waste, etc.. General-purpose steam sterilization of
surgical instruments, textiles and hospital utensils in central sterilization departments,
operation departments and laundries.
Competitive Advantage
Medpharm strong engineering capability allows for highly customized autoclave design.
Facilities with space layout limitations will benefit from custom space design autoclaves which
provide high capacity in a limited space.